タイトル |
|
|
タイトル |
Basic study of a diagnostic modality employing a new electrical impedance tomography (EIT) method for noninvasive measurement in localized tissue |
|
言語 |
en |
著者 |
岡﨑, 憲二
丹黒, 章
森本, 忠興
小谷, 亮介
服部, 圭悟
安野, 恵実子
芥川, 正武
木内, 陽介
|
抄録 |
|
|
内容記述タイプ |
Abstract |
|
内容記述 |
The objective of this study is to develop a device for noninvasive local tissue electrical impedance tomography (EIT) using divided electrodes with guard electrodes and to validate its effectiveness using bioequivalent phantoms. For this purpose, we prepared a measurement device and bioequivalent phantoms, measured the electrical characteristics of the phantoms, and validated the method using the phantoms. Monolayer phantoms mimicking the brain and muscle and bilayer phantoms consisting of muscle and brain layers were prepared. The relative differences between the measured electrical conductivities of the monolayer brain and muscle phantoms and the true values determined by the 4-electrode method were both less than 10%. The relative differences between the measured and true values in the bilayer phantoms were less than 20% in both layers. The biological impedance measurement device that we developed was confirmed to be effective for impedance measurement in bilayer phantoms with different electrical impedances. To develop a device for the early diagnosis of breast diseases, the development of a multi-layer phantom and demonstration of the effectiveness of the device for its examination are necessary. If the device that we developed makes impedance measurement in breast tumors possible, it may be used as a new diagnostic modality for breast diseases. |
|
言語 |
en |
キーワード |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
noninvasive |
キーワード |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
local tissue |
キーワード |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
electrical impedance tomography |
キーワード |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
measurement |
キーワード |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
breast diseases |
書誌情報 |
en : The journal of medical investigation : JMI
巻 57,
号 3-4,
p. 205-218,
発行日 2010-08
|
収録物ID |
|
|
収録物識別子タイプ |
ISSN |
|
収録物識別子 |
13431420 |
収録物ID |
|
|
収録物識別子タイプ |
NCID |
|
収録物識別子 |
AA11166929 |
備考 |
|
|
言語 |
ja |
|
値 |
The journal of medical investigation : http://medical.med.tokushima-u.ac.jp/jmi/index.html |
EID |
|
|
識別子 |
214721 |
|
識別子タイプ |
URI |
言語 |
|
|
言語 |
eng |