Noninvasive continuous blood pressure monitoring by the ClearSight system during robot-assisted laparoscopic radical prostatectomy

https://tokushima-u.repo.nii.ac.jp/records/2004598
https://tokushima-u.repo.nii.ac.jp/records/2004598
695f0f81-c574-4fb3-8737-1a87178ddb19
名前 / ファイル ライセンス アクション
jmi_65_1-2_69.pdf jmi_65_1-2_69.pdf (863 KB)
Item type 文献 / Documents(1)
公開日 2018-04-27
アクセス権
アクセス権 open access
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
出版社版DOI
関連識別子 https://doi.org/10.2152/jmi.65.69
関連名称 10.2152/jmi.65.69
出版タイプ
出版タイプ VoR
出版タイプResource http://purl.org/coar/version/c_970fb48d4fbd8a85
タイトル
タイトル Noninvasive continuous blood pressure monitoring by the ClearSight system during robot-assisted laparoscopic radical prostatectomy
著者 酒井, 陽子

× 酒井, 陽子

WEKO 656
徳島大学 教育研究者総覧 264524/profile-ja.html
e-Rad 90711862

ja 酒井, 陽子
ISNI

ja-Kana サカイ, ヨウコ

en Sakai, Yoko

Search repository
堤, 保夫

× 堤, 保夫

WEKO 343
e-Rad 90523499

ja 堤, 保夫
ISNI

ja-Kana ツツミ, ヤスオ

en Tsutsumi, Yasuo

Search repository
大山, 拓朗

× 大山, 拓朗

WEKO 843
徳島大学 教育研究者総覧 314552/profile-ja.html

ja 大山, 拓朗
ISNI

ja-Kana オオヤマ, タクロウ

en Oyama, Takuro

Search repository
村上, 千晶

× 村上, 千晶

WEKO 970
徳島大学 教育研究者総覧 350231/profile-ja.html
e-Rad 70825877

ja 村上, 千晶
ISNI

ja-Kana ムラカミ, チアキ

en Murakami, Chiaki

Search repository
角田, 奈美

× 角田, 奈美

WEKO 454
徳島大学 教育研究者総覧 208070/profile-ja.html
e-Rad 00622606

ja 角田, 奈美
ISNI

ja-Kana カクタ, ナミ

en Kakuta, Nami

Search repository
田中, 克哉

× 田中, 克哉

WEKO 1645
徳島大学 教育研究者総覧 60304/profile-ja.html
e-Rad 30263841

ja 田中, 克哉
ISNI

ja-Kana タナカ, カツヤ

en Tanaka, Katsuya

Search repository
抄録
内容記述 Robot-assisted laparoscopic radical prostatectomy (RALRP) is commonly performed in the surgical treatment of prostate cancer. However, the steep Trendelenburg position (25°) and pneumoperitoneum required for this procedure can sometimes cause hemodynamic changes. Although blood pressure is traditionally monitored invasively during RALRP, the ClearSight system (BMEYE, Amsterdam, The Netherlands) enables a totally noninvasive and simple continuous blood pressure and cardiac output monitoring based on finger arterial pressure pulse contour analysis. We therefore investigated whether noninvasive continuous arterial blood pressure measurements using the ClearSight system were comparable to those obtained invasively in patients undergoing RALRP. Ten patients scheduled for RALRP with American Society of Anesthesiologists physical status I-II were included in this study. At each of the seven defined time points, noninvasive and invasive blood pressure measurements were documented and compared in each patient using Bland-Altman analysis. Although the blood pressure measured with the ClearSight system correlated with that measured invasively, a large difference between the values obtained by the two devices was noted. The ClearSight system was unable to detect blood pressure accurately during RALRP, suggesting that blood pressure monitoring using this device alone is not feasible in this small patient population.
キーワード
主題 blood pressure
キーワード
主題 noninvasive
キーワード
主題 robot-assisted laparoscopic radical prostatectomy
書誌情報 en : The Journal of Medical Investigation

巻 65, 号 1-2, p. 69-73, 発行日 2018-02
収録物ID
収録物識別子タイプ ISSN
収録物識別子 13496867
収録物ID
収録物識別子タイプ ISSN
収録物識別子 13431420
収録物ID
収録物識別子タイプ NCID
収録物識別子 AA11166929
収録物ID
収録物識別子タイプ NCID
収録物識別子 AA12022913
出版者
出版者 Faculty of Medicine Tokushima University
EID
識別子 358033
言語
言語 eng
戻る
0
views
See details
Views

Versions

Ver.1 2024-10-28 03:28:16.069968
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3