タイトル |
|
|
タイトル |
Safety of the batteries and power units used in insulin pumps : A pilot cross‐sectional study by the Association for the Study of Innovative Diabetes Treatment in Japan |
タイトル別表記 |
|
|
その他のタイトル |
Safety of batteries in insulin pumps |
著者 |
ムラタ, タカシ
ニレンギ, シンスケ
サカネ, ナオキ
黒田, 暁生
ヒロタ, ユウシ
松久, 宗英
ナンバ, ミツヨシ
コバヤシ, テツロウ
|
抄録 |
|
|
内容記述 |
Aims/Introduction: We investigated the safety of the batteries and power units used in insulin pumps in Japan. Materials and Methods: A self‐administered questionnaire was sent to the 201 members of the Association for Innovative Diabetes Treatment in Japan. Results: A total of 56 members responded, and among the 1,499 active devices, 66 had episodes of trouble related to the batteries and power units. The ratio of reported troubles to the number of insulin pumps was significantly higher in insulin pumps with a continuous glucose monitoring sensor compared with insulin pumps without a continuous glucose monitoring sensor (odds ratio 2.82, P < 0.05). The cause and the consequences varied. The brands of the batteries varied; alkaline batteries purchased at drug stores and other shops accounted for 19.7%. Termination of battery life within 72 h of use was reported most frequently (50.0%), suspension of the insulin pump (21.2%) and leakage of the battery fluid (4.5%) followed. A total of 53.2% of the reported insulin pumps needed to be replaced, and 37.1% of them recovered after replacement of the battery. Conclusions: As trouble related to the batteries and power units of insulin pumps was frequent, practical guidance should be provided to respective patients regarding the use of reliable batteries, and to be well prepared for unexpected insulin pump failure. |
キーワード |
|
|
主題 |
Battery |
キーワード |
|
|
主題 |
Continuous subcutaneous insulin infusion |
キーワード |
|
|
主題 |
Insulin pump |
書誌情報 |
en : Journal of Diabetes Investigation
巻 9,
号 4,
p. 903-907,
発行日 2017-10-21
|
収録物ID |
|
|
収録物識別子タイプ |
ISSN |
|
収録物識別子 |
20401124 |
収録物ID |
|
|
収録物識別子タイプ |
NCID |
|
収録物識別子 |
AA12488319 |
出版者 |
|
|
出版者 |
Asian Association for the Study of Diabetes |
出版者 |
|
|
出版者 |
John Wiley & Sons |
権利情報 |
|
|
権利情報 |
This is an open access article under the terms of the Creative Commons Attribution‐NonCommercial‐NoDerivs License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made. |
EID |
|
|
識別子 |
340001 |
言語 |
|
|
言語 |
eng |