Possible Relationship between the Deteriorated Accuracy of Intermittent-Scanning Continuous Glucose Monitoring Device and the Contact Dermatitis : Post-hoc analysis of the ISCHIA Study
https://tokushima-u.repo.nii.ac.jp/records/2012172
https://tokushima-u.repo.nii.ac.jp/records/2012172dea75c0b-b6e0-4413-bf37-431442e745a9
名前 / ファイル | ライセンス | アクション |
---|---|---|
![]() |
|
Item type | 文献 / Documents(1) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
公開日 | 2024-08-01 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
アクセス権 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
アクセス権 | open access | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
資源タイプ | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
資源タイプ | journal article | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
出版タイプ | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
出版タイプ | VoR | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
タイトル | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
タイトル | Possible Relationship between the Deteriorated Accuracy of Intermittent-Scanning Continuous Glucose Monitoring Device and the Contact Dermatitis : Post-hoc analysis of the ISCHIA Study | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
タイトル別表記 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
その他のタイトル | Contact Dermatitis and the Accuracy of isCGM Device | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
著者 |
トヨダ, マサオ
× トヨダ, マサオ
× ムラタ, タカシ
× ヒロタ, ユウシ
× ホソダ, キミノリ
× カトウ, ケン
× コウヤマ, クンイチ
× コウヤマ, リュウジ
× 黒田, 暁生
WEKO
522
× マトバ, ユカ
× 松久, 宗英
WEKO
415
× メグロ, シュウ
× ミウラ, ジュンノスケ
× ニシムラ, クニヒロ
× シマダ, アキラ
× スズキ, ショウタ
× トネ, アツヒト
× サカネ, ナオキ
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
抄録 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
内容記述 | Objective: We previously reported the mean average relative difference (MARD) of the sensor glucose (SG) of the first-generation FreeStyle Libre with the original algorithm, an intermittent scanning continuous glucose monitoring (isCGM) device, was 15.6% in the Effect of Intermittent-Scanning Continuous Glucose Monitoring to Glycemic Control Including Hypoglycemia and Quality of Life of Patients with Type 1 Diabetes Mellitus Study (ISCHIA Study). In the present study, we aimed to further analyze its accuracy in detail by conducting a post-hoc analysis of the study. Methods: The ISCHIA Study was a multicenter, randomized, cross-over trial to assess the efficacy of isCGM. The SG levels of isCGM and the measured capillary blood glucose (BG) levels of 91 participants were used for the analysis. Results: Bland-Altman analysis showed bias of -13.0 mg/dl when the SG levels were compared to the BG levels, however no proportional bias was observed (r = 0.085). MARD of the participants without and with contact dermatitis were 15.0 ± 6.0% and 27.4 ± 21.4% (P = 0.001), respectively. Conclusion: There was negative bias in the SG levels of isCGM compared to the BG levels. There is a possibility that the complication of the contact dermatitis during isCGM use may be related with deteriorated accuracy of the SG levels. |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
キーワード | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
主題 | intermittent scanning continuous glucose monitoring | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
キーワード | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
主題 | accuracy | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
キーワード | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
主題 | type 1 diabetes | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
キーワード | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
主題 | contact dermatitis | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
書誌情報 |
en : The Tokai Journal of Experimental and Clinical Medicine 巻 48, 号 3, p. 83-90, 発行日 2023-09 |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
収録物ID | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
収録物識別子タイプ | ISSN | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
収録物識別子 | 21852243 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
出版者 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
出版者 | The Tokai Medical Association | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
EID | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
識別子 | 409200 | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
言語 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
言語 | eng |