Pullout Strength of Pedicle Screws Following Redirection After Lateral or Medial Wall Breach

https://tokushima-u.repo.nii.ac.jp/records/2007642
https://tokushima-u.repo.nii.ac.jp/records/2007642
a0e0857d-733b-4bf1-a725-da112ac503fc
名前 / ファイル ライセンス アクション
k3372_abstract_review.pdf k3372_abstract_review.pdf (176 KB)
k3372_fulltext.pdf k3372_fulltext.pdf (779 KB)
license.icon
アイテムタイプ 文献 / Documents(1)
公開日 2020-05-20
アクセス権
アクセス権 open access
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_db06
資源タイプ doctoral thesis
出版社版DOI
関連識別子 https://doi.org/10.1097/BRS.0000000000002611
関連名称 10.1097/BRS.0000000000002611
出版タイプ
出版タイプ NA
出版タイプResource http://purl.org/coar/version/c_be7fb7dd8ff6fe43
タイトル
タイトル Pullout Strength of Pedicle Screws Following Redirection After Lateral or Medial Wall Breach
タイトル別表記
その他のタイトル 椎弓根スクリューの椎弓根外側、内側に誤挿入後再挿入した際の引き抜き強度
タイトル別表記
その他のタイトル Pullout Strength of Pedicle Screws
著者 前田, 徹

× 前田, 徹

ja 前田, 徹

ja-Kana マエダ, トオル

en Maeda, Toru

Search repository
東野, 恒作

× 東野, 恒作

WEKO 182
e-Rad_Researcher 80380129

ja 東野, 恒作
東野, 恒作
ISNI

ja-Kana ヒガシノ, コウサク
ヒガシノ, コウサク

en Higashino, Kousaku
Higashino, Kousaku

Search repository
眞鍋, 裕昭

× 眞鍋, 裕昭

WEKO 941
徳島大学 教育研究者総覧 346406/profile-ja.html

ja 眞鍋, 裕昭
ISNI

ja-Kana マナベ, ヒロアキ

en Manabe, Hiroaki

Search repository
山下, 一太

× 山下, 一太

WEKO 904

ja 山下, 一太
ISNI

ja-Kana ヤマシタ, カズタ

en Yamashita, Kazuta

Search repository
林, 二三男

× 林, 二三男

ja 林, 二三男

ja-Kana ハヤシ, フミオ

en Hayashi, Fumio

Search repository
合田, 有一郎

× 合田, 有一郎

ja 合田, 有一郎

ja-Kana ゴウダ, ユウイチロウ

en Goda, Yuichiro

Search repository
鶴尾, 吉宏

× 鶴尾, 吉宏

WEKO 755
徳島大学 教育研究者総覧 289950/profile-ja.html
e-Rad_Researcher 90207449

ja 鶴尾, 吉宏
鶴尾, 吉宏
ISNI

ja-Kana ツルオ, ヨシヒロ
ツルオ, ヨシヒロ

en Tsuruo, Yoshihiro
Tsuruo, Yoshihiro

Search repository
西良, 浩一

× 西良, 浩一

WEKO 1543
徳島大学 教育研究者総覧 60063/profile-ja.html
e-Rad_Researcher 10304528

ja 西良, 浩一
西良, 浩一
ISNI

ja-Kana サイリョウ, コウイチ
サイリョウ, コウイチ

en Sairyo, Koichi
Sairyo, Koichi

Search repository
抄録
内容記述 Study Design: A cadaveric biomechanical study designed to test the pullout strength of pedicle screws.
Objective: To evaluate the pullout strength of redirected pedicle screws with a larger diameter following lateral wall breach, redirected pedicle screws of the same diameter following medial wall breach, and redirected pedicle screws with a larger diameter following medial wall breach. Summary of Background Data: Screw malposition is one of the main pitfalls of inserting pedicle screws. Intraoperatively a malpositioned screw is re-directed and inserted along the correct axis.
Methods: Forty-seven vertebrae (T9–L5) were harvested from 8 fresh cadaveric spines. The 18 pedicle screws that breached the lateral wall were then removed and redirected using a pedicle screw of 1 mm larger in diameter. The 16 pedicle screws that had breached the medial wall were then removed and redirected using a pedicle screw of the same diameter. The other 13 pedicle screws that had breached the medial wall were then removed and redirected using a pedicle screw of 1 mm larger in diameter. The pullout strength was measured.
Results: Following lateral wall breach, mean pullout strength for the larger redirected screws was 46.9% greater than that of the correctly aligned screws. Following medial wall breach, mean pullout strength for the redirected screws of the same diameter was 20.6% less than that of the correctly aligned screws. Mean pullout strength for the larger pedicle screws following medial wall breach was 27.3% more than that of the correctly aligned screws.
Conclusion: Redirected pedicle screws of larger diameter after a lateral or medial pedicle breach show recovery of pullout strength. However, the pullout strength of redirected pedicle screws of the same diameter after a medial pedicle breach is significantly less than that of correctly aligned screws.
抄録
内容記述 Redirected pedicle screws of larger diameter after a lateral or medial pedicle breach show the recovery of pullout strength. However, the pullout strength of redirected pedicle screws of the same diameter after a medial pedicle breach is significantly less than that of correctly aligned screws.
キーワード
主題 larger screw
キーワード
主題 malpositioned screw
キーワード
主題 pedicle screw
キーワード
主題 pullout strength
キーワード
主題 redirection
キーワード
主題 thoracolumbar spine
書誌情報 en : Spine

巻 43, 号 17, p. E983-E989, 発行日 2018-09-01
収録物ID
収録物識別子タイプ ISSN
収録物識別子 03622436
収録物ID
収録物識別子タイプ ISSN
収録物識別子 15281159
収録物ID
収録物識別子タイプ NCID
収録物識別子 AA0084801X
収録物ID
収録物識別子タイプ NCID
収録物識別子 AA12431865
出版者
出版者 Lippincott Williams & Wilkins
出版者
出版者 Wolters Kluwer Health, Inc.
備考
値 内容要旨・審査要旨・論文本文の公開
本論文は,著者Toru Maedaの学位論文として提出され,学位審査・授与の対象となっている。
学位授与者所属 : 徳島大学大学院医科学教育部(医学専攻)
権利情報
権利情報 This the final peer-reviewed manuscript is made available under the CC-BY-NC 4.0 license.
EID
識別子 340874
言語
言語 eng
報告番号
学位授与番号 甲第3372号
学位記番号
値 甲医第1455号
学位授与年月日
学位授与年月日 2020-03-23
学位名
学位名 博士(医学)
学位授与機関
学位授与機関名 徳島大学
戻る
0
views
See details
Views

Versions

Ver.1 2024-10-30 09:41:41.303623
Show All versions

Share

Share
tweet

Cite as

Other

print

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX
  • ZIP

コミュニティ

確認

確認

確認


Powered by WEKO3


Powered by WEKO3