当院におけるカプセル内視鏡検査の画像不良事例と対策について
https://tokushima-u.repo.nii.ac.jp/records/2008294
https://tokushima-u.repo.nii.ac.jp/records/20082947f133c3b-932f-4cc6-98ff-a397e9f4e5c2
名前 / ファイル | ライセンス | アクション |
---|---|---|
sam_76_3-4_159.pdf (1.58 MB)
|
|
Item type | 文献 / Documents(1) | |||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
公開日 | 2020-10-20 | |||||||||||||||||||||||
アクセス権 | ||||||||||||||||||||||||
アクセス権 | open access | |||||||||||||||||||||||
資源タイプ | ||||||||||||||||||||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||||||||||||||||||||
資源タイプ | journal article | |||||||||||||||||||||||
出版タイプ | ||||||||||||||||||||||||
出版タイプ | VoR | |||||||||||||||||||||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 | |||||||||||||||||||||||
タイトル | ||||||||||||||||||||||||
タイトル | 当院におけるカプセル内視鏡検査の画像不良事例と対策について | |||||||||||||||||||||||
言語 | ja | |||||||||||||||||||||||
タイトル別表記 | ||||||||||||||||||||||||
その他のタイトル | Assessment of accidental events in images during capsule endoscopy | |||||||||||||||||||||||
言語 | en | |||||||||||||||||||||||
タイトル別表記 | ||||||||||||||||||||||||
その他のタイトル | カプセル内視鏡における画像不良 | |||||||||||||||||||||||
言語 | ja | |||||||||||||||||||||||
著者 |
藤井, 有美子
× 藤井, 有美子
× 小林, 誠司
× 大西, 芳明
× 田中, 克哉× 六車, 直樹× 高山, 哲治 |
|||||||||||||||||||||||
抄録 | ||||||||||||||||||||||||
内容記述タイプ | Abstract | |||||||||||||||||||||||
内容記述 | Recently, there have been some reports of electromagnetic interference during capsule endoscopy(CE)in regards to other medical devices and facility environment. These accidental events have not been analyzed sufficiently although academic societies and related ministries are proposing guidelines with cautious attention. Based on this, detailed analyses and development of solution are required. In this study, we review our experience with cases of accidental events during CE and assessed the causes and preventive solutions. A total of149CE cases(male :89, female :60, mean age :56±18.8)from January 2017 until December 2019 were analyzed retrospectively. Four cases(2%)with defective images were noted. Detailed events were as follows ;1 disorder of the sensor array, 2 recorder freezes, and 1 electromagnetic interference. These mechanical and electromagnetic troubles should be assessed and shared among medical staffs and manufacturers to propose possible solutions and perform CE safely. | |||||||||||||||||||||||
言語 | en | |||||||||||||||||||||||
キーワード | ||||||||||||||||||||||||
言語 | en | |||||||||||||||||||||||
主題Scheme | Other | |||||||||||||||||||||||
主題 | Capsule endoscopy | |||||||||||||||||||||||
キーワード | ||||||||||||||||||||||||
言語 | en | |||||||||||||||||||||||
主題Scheme | Other | |||||||||||||||||||||||
主題 | Image defect | |||||||||||||||||||||||
キーワード | ||||||||||||||||||||||||
言語 | en | |||||||||||||||||||||||
主題Scheme | Other | |||||||||||||||||||||||
主題 | Electromagnetic interference | |||||||||||||||||||||||
書誌情報 |
ja : 四国医学雑誌 en : Shikoku Acta Medica 巻 76, 号 3-4, p. 159-164, 発行日 2020-08-25 |
|||||||||||||||||||||||
収録物ID | ||||||||||||||||||||||||
収録物識別子タイプ | ISSN | |||||||||||||||||||||||
収録物識別子 | 00373699 | |||||||||||||||||||||||
収録物ID | ||||||||||||||||||||||||
収録物識別子タイプ | NCID | |||||||||||||||||||||||
収録物識別子 | AN00102041 | |||||||||||||||||||||||
出版者 | ||||||||||||||||||||||||
出版者 | 徳島医学会 | |||||||||||||||||||||||
言語 | ja | |||||||||||||||||||||||
言語 | ||||||||||||||||||||||||
言語 | jpn |